Medcial Equipemt Chemistry Analyzer Fully Automatic Urint Semiautomated Chemistry Analyzer BS-200


Performance
Dynamic and real time display of running status
.Running status of reagent tray,
sample tray and reaction tray
Realtime monitoring of reagent residual volume
Intelligent carry over settings to adjust test sequence
Probe depth adjustment automatically
Original reaction data record
.Real time monitoring of reaction curve
.Bichromatic testing to avoid interference Simultaneously display primary and secondary wavelengths
.Detailed profile of alert messages
.Realtime diagnosis of system working status
.Bichromatic testing to avoid interference Simultaneously display primary and secondary wavelengths
.Detailed profile of alert messages
.Realtime diagnosis of system working status
Optimum calibration curve
.Calibration classification: linear curve type: Single-point linear,two-point linear and Mult point linear.
.Nonlinear curve type: Logistic
Log 4P,Logistic-Log 5P,Exponential5P ,Polynomial5P and Spline
.Nonlinear curve type: Logistic
Log 4P,Logistic-Log 5P,Exponential5P ,Polynomial5P and Spline
Technical Specifications | |||
Throughput | up to 200 tests/hour (without ISE). up to 330 tests/hour with ISE | ||
Internal Bar Code Reader (optional): | Used for sample and reagent programming; applicable to various bar code systems including Codabar,ITF,Code128, Code 39,UPC / EAN,Code 93. Bi-directional LIS interface | ||
Sample Volume | 2~45μl, step by 0.1μl | ||
Measuring principles | Absorbance photometry, Turbidimetry Absorbance photometry Turbidimet Ion Selective Electrode technology | ||
Sample tray | 40 positions for reagents and 40 positions for samples in refrigerated compartment (4-15 ºC) | ||
Reagent Handling: | Reagen t/Sample tray: 40 positions for reagents and 40 positions for samples in refrigerated compa ment (4-15 ºC) Reagent volume R1: 180-450IJI ,step by 1μ| R2: 30-450ul ,step by 1μl Sample volume: 3-45ul, step by 0.5ul Reagen t/Sample probe: liquid level detection ,collision protection and inventory checking Probe cleaning: automatic washing for both inside and outside carry over < 0.1 % Automatic sample dilution: pre-dilution and post-dilution dilution ratio up to 150 dilution vessel: disposable cuvette | ||
Sample cup type | Micro cup & test tube | ||
Reaction System | Reaction rotor: rotating tray ,containing 80 cuvettes Cuvette: optical length 5mm Reaction volume: 180-500 IJI Reaction temperature: 37 C Temperature fluctuation: 0.1 C Independent mixing probe | ||
Operation system | Windows® XP Professional/Home SP2 or above Windows® Vista Home/Business, Windows® 7 | ||
Interface | RS-232 | ||
Working Conditions | Power Supply: AC200-240V ,50/60Hz ,1000W or AC100-130V,50/60Hz.1000W Temperature:15-30 C Humidity:35-80% Water consumption 3.5υhour De-ionized | ||
Dimension | Bench top: 860mm(W)x700 mm(D)x625 mm (H) Floor standing: 860mm(W)x700 mm(D)x1160 mm (H) | ||
Weight | Bench top: 116 kg Cabinet (optional): 51 kg |